Services details

ELECTROLYTE ANALYZER

  • Na+, K+, CI- /Ca++
  • Low Sample Volume
  • Fast Sample Processing
  • Low Cost per Test

Technical Specification

Principle Ion selective Electrode (ISE)
Sample Types Serum, Plasma, Whole Blood, Urine
Parameters Na+, K+, CI / Ca++
Sample volume s 90 uL
Analysis Time s 60 sec
Throughput s 60 samples per hour
Calibration Automatic Calibration Two Point Calibration after 4 hrs or just before Sample Analysis (whichever is longer) One point Calibration with each Sample Analysis
Normal Ranges User defined normal ranges for all sample types with flagging for Abnormal samples
Data Management Calibration History, Sample History and Control History of last 30 days
Quality Control Tri-LevelQuality Control. with LJ Plots of last 30 days
Reagent Pack Convenient, easy to place ISEPAK which contains reagents required for all Electrolyte Parameters
Reagent status Tracks reagents consumption and provides alerts for low reagent volume
Display 5" color LCD with Touchscreen Display, Resolution: 4B0 x 272 pixels
Printer Built-in 2" (48 mm) Thermal Printer, 42 characters wide
External Interface USB (Keyboard, Mouse)
Operating Temp. 15 to 35 C
Relative Humidity 5 to 85% non-condensing
Power Supply 100 to 240 VAC 50/60 Hz
Dimensions 395 mm x 260 rnm x 375 mm
Weight | Sample Storage 7.5 Kg | 1100,000